ION930 Benchtop pH/Ion Meter


Product Description

ION930 Benchtop pH/Ion Meter 


Features:

1. Automatic Temperature Compensation ensures accurate reading over the entire range.

2. Calibration Due Reminder prompts user to calibrate the meter regularly.

3. Stability indicator automatically shows current measurement status.

4. Auto-Hold function freezes stable measured value for easy viewing and recording.

5. Manual temperature calibration improves the measurement accuracy of meter.

6. Selectable temperature unit (°C or °F) meets different application requirements.

7. Automatic electrode diagnosis shows zero point offset and slope of the electrode.

8. Help message as a operational guide that helps you quickly begin using the meter.

9. System menu allows the setting 10 parameters, including the pH buffer standard, number of calibration points, resolution, stability condition, hold function, etc.

10. Reset feature automatically resumes all parameters back to factory default settings.

11. Expanded memory stores and recalls up to 500 readings.

12. Built-in real-time clock stamps stored data to meet GLP standard.

13. Stored data can be transferred into computer by USB communication interface.


Specifications:
pH
1. Range: -2.000~20.000pH
2. Accuracy: ±0.002pH
3. Resolution: 0.1, 0.01, 0.001pH, selectable
4. Calibration Points: 1 to 5 points
5. pH Buffer Options: USA, NIST, DIN or Custom
 
mV
1. Range: ±1999.9mV
2. Accuracy: ±0.2mV
3. Resolution: 0.1, 1mV, selectable
4. Calibration Points: 1 point
5. Measurement Modes: Relative or Absolute mV
 
Ion
1. Range: 0.001~19999 (Deponding on the range of ISE)
2. Accuracy: ±0.5% F.S (Monovalent), ±1% F.S (Divalent)
3. Resolution: 0.001, 0.01, 0.1, 1
4. Calibration Points: 2 to 5 points (0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000)
5. Measurement Units: ppm, mg/L, mol/L, mmol/L
 
Temperature: 
1. Range: 0~105°C, 32~221°F
2. Accuracy: ±0.5°C
3. Resolution: 0.1°C, 0.1°F
4. Calibration Points: 1 point
5. Calibration Range: Measured value ±10°C
 
General
1. Temperature Compensation: 0~100°C, 32~212°F, Manual or Automatic
2. Data Hold: Manual or Auto-Endpoint
3. Stability Criteria: Low or High
4. Calibration Due Alarm: 1 to 31 days or Off
5. Reset Function: Yes
6. Memory: Stores up to 500 data sets
7. Output: USB communication interface
8. Connector: BNC
9. Display: LCD
10. Power: DC5V, using AC adapters, 220VAC/50Hz
11. Dimensions: 210x188x60mm
12. Weight: 1.5kg